- Braided rigid shaft for optimal support and pushability
- Tungsten tip and markers for high visibility
- Low profile tip and smooth taper ideal for telescoping technique
Tungsten-compounded resin provides visibility to the most distal end of the tip and at known marker bands (30mm and 50mm) on the shaft.
CROSSWALK tip, 3cm and 5 cm
radiopaque markers are at the same
distance on all size catheters.
CROSSWALK 0.018" shown is an angled type catheter.
Catheter | Tip entry OD | Tip entry ID | Distal crossing OD |
Proximal shaft OD |
---|---|---|---|---|
0.014" | 0.53mm (1.6Fr.) |
0.38mm (0.015") |
0.72mm (0.029"/2.2F) |
0.89mm (0.035"/2.7F) |
0.018" | 0.59mm (1.7Fr.) |
0.48mm (0.019") |
0.84mm (0.033"/2.5F) |
0.89mm (0.035”/2.7F) |
0.035" | 1.07mm (3.2Fr.) |
0.93mm (0.037") |
1.40mm (0.055"/4.2F) |
1.40mm (0.055"/4.2F) |
Thin braided shaft reinforced by outer resin delivers high pushability and torque control while maintaining low profile.
CROSSWALK microcatheter delivers high shaft stiffness with
smooth transitions for optimal support profile.
CROSSWALK
Other Competitive Devices
PROXIMAL SHAFT STIFFNESS
Test Method:
Evaluation of the proximal
shaft stiffness by using
three-point bend test.
* All the data were obtained by company standardized test, which may differ from industry standardized tests.
* All the data do not guarantee that all devices have exactly the same performance with the samples used for tests.
Deliverability and crossability are critical to treatment success. CROSSWALK's stiff shaft and telescoping technique enhances deliverability across highly stenosed lesions.
CROSSWALK's telescope technique enables a 0.014" or 0.018" catheter to fit inside an 0.035" catheter. This increases shaft support and pushability to facilitate crossing of highly stenosed lesions.
Highly stenosed lesions may cause the catheter shaft to deflect. As a result, push force is lost and the support catheter may have difficulty crossing the lesion.
No 90cm marker in 90cm length product.
Product | Catalog No. |
Recommended guide wire |
Usable length |
Tip shape |
Outer diameter | Inner diameter | |||
---|---|---|---|---|---|---|---|---|---|
Tip | Crossing profile |
Proximal | Tip | Shaft | |||||
CROSSWALK | PSC14090S | 0.36mm (0.014") |
90cm | Straight | 0.53mm (1.6Fr.) |
0.72mm, 0.029" (2.2Fr.) |
0.89mm, 0.035" (2.7Fr.) |
0.38mm (0.015") |
0.42mm (0.016") |
PSC14110S | 110cm | ||||||||
PSC14135S | 135cm | ||||||||
PSC14150S | 150cm | ||||||||
PSC14170S | 170cm | ||||||||
PSC14090A | 90cm | Angled |
|||||||
PSC14110A | 110cm | ||||||||
PSC14135A | 135cm | ||||||||
PSC14150A | 150cm | ||||||||
PSC14170A | 170cm | ||||||||
PSC18090S | 0.46mm (0.018") |
90cm | Straight | 0.59mm (1.7Fr.) |
0.84mm, 0.033" (2.5Fr.) |
0.89mm, 0.035" (2.7Fr.) |
0.48mm (0.019") |
0.53mm (0.021") |
|
PSC18110S | 110cm | ||||||||
PSC18135S | 135cm | ||||||||
PSC18150S | 150cm | ||||||||
PSC18170S | 170cm | ||||||||
PSC18090A | 90cm | Angled |
|||||||
PSC18110A | 110cm | ||||||||
PSC18135A | 135cm | ||||||||
PSC18150A | 150cm | ||||||||
PSC18170A | 170cm | ||||||||
PSC35090S | 0.89mm (0.035") |
90cm | Straight | 1.07mm (3.2Fr.) |
1.40mm, 0.055" (4.2Fr.) |
1.40mm, 0.055" (4.2Fr.) |
0.93mm (0.037") |
0.99mm (0.039") |
|
PSC35110S | 110cm | ||||||||
PSC35135S | 135cm | ||||||||
PSC35150S | 150cm | ||||||||
PSC35170S | 170cm | ||||||||
PSC35090A | 90cm | Angled |
|||||||
PSC35110A | 110cm | ||||||||
PSC35135A | 135cm | ||||||||
PSC35150A | 150cm | ||||||||
PSC35170A | 170cm |